FDAFebruary 25, 2022device

syngo Application software VE20 (Material Number 10848815) installed on the Artis pheno and Artis icono systems as follows: (1) Artis pheno Model Number:10849000 (2) Artis icono biplane Model Number: 11327600 (3) Artis icono floor Model Number: 11327700

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

After CT image data from Toshiba is loaded, image mirroring can occur along the horizontal and vertical image axes. If this error occurs, the patient orientation/position may be misinterpreted and result in inappropriate treatment, even if the incorrect visualization is obvious.

What to do

FDA enforcement status: Ongoing

Brands named

siemens medical solutionssiemenssiemens medical