FDADecember 12, 2025device

Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system Software Version: 11.1 Product Description: Leksell GammaPlan¿ is a computer-based system designed for Leksell Gamma Knife¿ treatment planning.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Due to an error when defining a new stereotactic reference that users fail to first change an already defined obsolete reference to a pre-plan reference. This may result in a treatment plan that potentially irradiates the incorrect patient location if the error is not detected by users.

What to do

FDA enforcement status: Ongoing

Brands named

elekta

UPCs

07340048311236

Recall history