FDAJanuary 31, 2020device

UST-2265-2 Probe, Part Number: UST-2265-2 - Product Usage: The UST-2265-2 probe is intended to be used for diagnostic ultrasound imaging or fluid flow analysis of the human body as Continuous-wave Doppler (CWD) for Cardiac Adult, Cardiac Pediatric, and Peripheral Vessel.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

These 18 probes may have mis-wiring in the inner cables, causing decreased sensitivity and thus resulting in lower peak velocity readings.

What to do

FDA enforcement status: Terminated

Brands named

hitachi medical systems americahitachihitachi medical