FDAMay 20, 2022device

CELLSEARCH Circulating Tumor Cell Kit (Epithelial)-IVD intended for the enumeration of circulating tumor cells (CTC) of epithelial origin (CD45-, EpCAM+, and cytokeratins 8, 18+, and/or 19+) in whole blood Part Number: 7900001

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

High number of total images/unassigned events including (dual positives) and the potential for false positive results being placed into the image gallery in some patient samples

What to do

FDA enforcement status: Ongoing

Brands named

menarini silicon biosystemsmenarinimenarini silicon