FDAMarch 24, 2023device

Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff, and Cuffless: Disposable Inner Cannula or Reusable Inner Cannula (REF, Product Description): 10CN10H, 10.0MM SHILEY CUFFED TRACH CAN; 10CN10R, 10.0MM ADT FLEX TRACH W TG CUFF; 10UN10A, 10.0MM UNCUFF TRACH TUBE X1; 10UN10H, 10.0MM SHILE...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

A manufacturing error, resulted in a less than specified diameter of adult flexible tracheostomy tube connectors, which results in unsecure connections with 15mm caps and other 15mm circuit components and accessories. An unsecure connection could lead to respiratory failure, dyspnea, treatment delay, tissue injury, or bleeding.

What to do

FDA enforcement status: Ongoing

Brands named

covidien

UPCs

108845211723711088452120547520884521175690208845211762911088452120554310884521172456208845211763071088452120502420884521175706208845211762391088452120548210884521172463

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff, and Cuffless: Disposable Inner Cannula or Reusable Inner Cannula (REF, Product Description): 10CN10H, 10.0MM SHILEY CUFFED TRACH CAN; 10CN10R, 10.0MM ADT FLEX TRACH W TG CUFF; 10UN10A, 10.0MM UNCUFF TRACH TUBE X1; 10UN10H, 10.0MM SHILE... — Recall Details · AllClear