FDAFebruary 20, 2015device

GAP Nail 4.8mm (diameter size) of the GAP ENDO-EXO MEDULLARY SYSTEM. Catalog number GAP-N48-16; GAP-N48-18; GAP-N48-26; GAP-N48-28. Product Usage: The GAP Endo-Exo Medullary System is used for the treatment of fractures or correction of deformities in the femur, tibia and humerus of pediatric pat...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Failures of the 4.8mm of the GAP Endo-Exo Medullary System have been identified at the level of the distal cortical screw holes. Failures mainly occurred in patients with tibial pseudoarthrosis (CPT) or Osteogenesis imperfecta (OI). Higher failure rates have been detected in tibias with distal fractures that are close to the distal locking holes.

What to do

FDA enforcement status: Terminated

Brands named

pega medicalpega

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.