FDAJanuary 10, 2020device
NUCLISENS¿ Lysis Buffer, IVD, REF 200292, 48X2 ML LYS Manufacturer bioMerieux SA 376, Chemin de l'Orme 69280 Marcy-l'Etoile - France - Product Usage: to be used for the release of total nucleic acid from biological specimens.
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Following a customer complaint about colored eluates, leading to invalid results bioMerieux has confirmed an issue with NUCLISENS Lysis Buffer, ref. 200292, lot 19040201. The issue is linked to drift of pH electrode dysfunction.
What to do
FDA enforcement status: Terminated
Brands named
biomerieux grenoblebiomerieuxbiomerieux grenoble
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAVITEK 2 Gram-negative Susceptibility Cards Containing o Polymyxin B (pb02n) o Ceftazidime/avibactam (cza02n) o Meropenem/Vaborbactam (mev01n) o Imipenem (ipm05n) o Imipenem/Relebactam (ipr01n); CARD NAME/ PRODUCT REFERENCE: AST-N399 423701 AST-N402 423644 AST-N403 423645 AST-N404 423664 AST-N405 ...2026-01-05
- FDAVITEK¿ 2 Gram-negative AST Test Kits (Cards) containing Colistin formulation cs02n. VITEK¿ 2 is an automated system consisting of instruments, software and reagent cards designed for the identification and antimicrobial susceptibility testing of bacteria and yeast. Export only2025-06-26
- FDAVITEK 2 - AST Test kits (in vitro diagnostic test cards for antimicrobial susceptibility): a) AST-N334, REF 418984; b) AST-GN95, REF 421982; c) AST-GN99, REF 423102; d) AST-N390, REF 423340; e) AST-N391, REF 423341; f) AST-N392, REF 423342; g) AST-N395, REF 423491; h) AST-N401, REF 423643; i) AST...2024-03-14
- FDAMYLA comprises AST Filters in conjunction with VITEK MS; middleware which connects microbiology laboratory instruments for ID/AST and blood culture activities and the laboratory information system (LIS).2023-02-06
- FDAVITEK 2 REAGENT GN TEST KIT VTK2 20 CARDS, CATALOG 213412022-11-09
- FDAVITEK 2 REAGENT GP TEST KIT VTK2 20 CARDS, CATALOG 213422022-11-09
- FDATEMPO REAGENTS TEMPO CONTROL KIT , CATALOG 800002022-11-09
- FDAVIDAS CLINICALVIDAS VARICEL. ZOSTER IGG 60 T, CATALOG 302172022-11-09
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →