GE Healthcare, Aespire 7900, Aespire View Anesthesia Machines. Model numbers 1009-9012-000. Intended to provide general inhalation anesthesia and ventilatory support to a wide range of patients

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Specific Aespire 7900 Anesthesia Machines have an increased likelihood of failure of the CPU circuit board. This could result in a blank display or loss of mechanical ventilation.

What to do

FDA enforcement status: Terminated

Brands named

ge healthcare

UPCs

10099012000