FDAJanuary 5, 2021device

HAMILTON-H900 Humidifier, Models: 950001, 950004

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

When starting a humidifier, intended for respiratory gas conditioning during invasive and noninvasive mechanical ventilation, with an empty chamber, and water is not filled into the chamber, the low water level alarm will be delayed and medical staff are unaware that there is no water in the chamber.

What to do

FDA enforcement status: Completed

Brands named

hamilton medical aghamiltonhamilton medical