FDAJanuary 7, 2026device
Medline Convenience Kits: 1) OR ARTERIOGRAM, Model Number: CDS985094F; 2) KIT SURG ONC LAP DIAG/CV CATH, Model Number: DYKMBNDL153; 3) KIT ACES INSERT PERCUTANEOUS T, Model Number: DYKMBNDL4A; 4) KIT ACES INSERT PERCUTANEOUS T, Model Number: DYKMBNDL4B; 5) VEIN PROCEDURE KIT, Model Number: DYNDA2...
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
What to do
FDA enforcement status: Ongoing
Brands named
medline industries lpmedlinemedline industries
UPCs
101953276576424019532765764310193489478440401934894784411019348990451240193489904513101953272005894019532720058010198459001208401984590012091019845910642240198459106423
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDAMedline Convenience Kits: 1) DRAWER 6 CENTRAL LINE INSERTIO, Model Number: ACC011142; 2) CENTRAL AND ARTERIAL LINE TOTE, Model Number: DYNJ84072026-05-06
- FDAMedline Convenience Kits: 1) DRAWER 6A ADULT CENTRAL LINE, Model Number: ACC010890; 2) OPEN HEART A OVERHEAD, Model Number: DYNJ910066; 3) OPEN HEART, Model Number: DYNJ9855016V; 4) OPEN HEART, Model Number: DYNJ9855016X2026-05-06
- FDAMedline Convenience Kits: 1) OPEN HEART CDS, Model Number: CDS840144Q; 2) OPEN HEART, Model Number: CDS840150W; 3) OPEN HEART ADULT, Model Number: CDS840396AA; 4) CABG A CDS, Model Number: CDS982269N; 5) OPEN HEART ANESTHESIA PACK, Model Number: CDS983849G; 6) OPEN HEART ACCESSORY CDS, Model Numb...2026-05-06
- FDAMedline Convenience Kits: 1) DYNJ905503F, Model Number: CV ANESTHESIA - ROOM SET UP; 2) DYNJ905503G, Model Number: CV ANESTHESIA - ROOM SET UP2026-05-06
- FDAMEDLINE Urethral Catheter, Red Rubber Latex, Medline Product Number/SKU (REF), French sizes: 1. DYND13508 - 08 Fr; 2. DYND13510 - 10 Fr; 3. DYND13512 - 12 Fr; 4. DYND13514 - 14 Fr; 5. DYND13515 - 15 Fr; 6. DYND13516 - 16 Fr; 7. DYND13518 - 18 Fr; 8. DYND13520 - 20 Fr; 9. DYND13522 - 22 Fr; 10. DY...2026-05-04
- FDAMEDLINE Medical Procedure Kits labeled as: PELVIC LAP PACK, Medline Kit Number/ SKU DYNJ53820B2026-05-04
- FDAMEDLINE Medical Procedure Kits labeled as: LAPAROSCOPY COMBINED - NO NDL, Medline Kit Number/SKU DYNJ37378A2026-05-04
- FDAMEDLINE Medical Procedure Kits labeled as: TONSIL PACK CATH, Medline Kit Number/SKU DYNJ65199A2026-05-04
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →