FDAFebruary 20, 2024device

Cardiovascular Procedure Kits that include High Crack Check Valve PN 205307A Part Number: 6375 16010 62974 65246 66312 66342 66388 66650 66852 70472 74307 74365 74366 74430 74475 74477 74502 74773 74873 75065 75157 75195 75409 75410 75473 75533 75534 75607 75765 75797 75833 75903 75956 75973 7613...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Pressure relief valve included in certain lots of cardiovascular procedure kits was reportedly opening at approximately 200-300mmHg versus 400mmHg, potential myocardial tissue damage potentially requiring surgical or medical/pharmacological intervention

What to do

FDA enforcement status: Ongoing

Brands named

terumo cardiovascular systemsterumoterumo cardiovascular

UPCs

00699753506657

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →