FDAMarch 5, 2021device

Covidien Surgiwand" II Suction and Irrigation Device: Electrosurgical, cutting & coagulation & accessories22 Product Number/CFN: 178083, 178093, 178094

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Investigations related to customer complaints received for particulate matter inside of the Surgiwand" II device tubing components identified the root cause as a damaged Y-Connector component which connects the tubing to the device. This damage was determined to generate the particulate matter observed and is created during the assembly process for connecting the tubing component to the Y-Connector.

What to do

FDA enforcement status: Ongoing

Brands named

covidien

UPCs

108845230008872088452300088410884523000900208845230009071088452300091720884523000914