FDAJanuary 10, 2020device

AIA-360 Automated Immunoassay Analyzer, Product Code 019945

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

A display screen software issue on the analyzer causes the display to freeze when display screen is touched at the same time as a command from the instrument firmware, causing the instrument to stop. As a result, the run is aborted and the results are not retrievable.

What to do

FDA enforcement status: Terminated

Brands named

tosoh biosciencetosoh