FDAMarch 14, 2023device

HAMILTON-C6, REF: 160021

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Software error causes, safety ventilation, in which ventilation continues in the "safety ventilation" mode with audible/visible alarm - patient inputs are not monitored, if the following coincide 1) A mode change to an adaptive mode (ASV, APVcmv, APVsimv, INTELLiVENT-ASV, (S)CMV+, SIMV+), and 2) The controller and/or humidifier is connected to the ventilator and is operational.

What to do

FDA enforcement status: Ongoing