FDAFebruary 26, 2021device

Instructions for Use and Patient Manual for HeartWare HVAD System. Accompanies these devices: Heartware HVAD pump implant kits Models 1100, 1101, 1102, 1103, 1104, 1104JP, 1205, and MCS1705PU. Heartware HVAD controller kits Models 1403US, 1407AU, 1407CA, 1407CH, 1407DE, 1407GB, 1407IL, 1407IN, 14...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

IFU and Patient Manual updated to describe HVAD Controller Connection Verification. The labeling is being corrected to reflect the actual verified behavior of a successful connector to the controller, which is a "beep" and not "audible click", to clarify under what alarm conditions a beep will not be heard, and identify alternate confirmation methods (visual indicator) for successful connection. NOTE: this update is included in the February 2021 Urgent Medical Device Notice also covered in recalls 87456 and 87603

What to do

FDA enforcement status: Ongoing

Brands named

heartware

UPCs

008887070000170088870700069700888707005364008887070004750088870700166300888707002851008887070016940088870700170000888707001717008887070017240088870700173100888707001748

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.