FDAMarch 18, 2015device

Magellan Robotic System; Model No(s). 11132 and 11139. A steerable catheter control system used at clinical sites.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Based on investigation of two complaint incidents, all Hansen Medical Magellan Robotic System (Model No. 11132 and 11139) may experience an unexpected event during a retraction operation of catheter. Retraction of catheter may not stop when equipment button is released.

What to do

FDA enforcement status: Terminated

Brands named

hansen medicalhansen

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →