FDAFebruary 20, 2020device

RayStation/RayPlan 8B, Build Number 8.1.0.47, stand-alone software treatment planning system

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the lateral, longitudinal and vertical values propagated to be incorrect.

What to do

FDA enforcement status: Terminated

Brands named

raysearch laboratories abraysearchraysearch laboratories

UPCs

07350002010129

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
RayStation/RayPlan 8B, Build Number 8.1.0.47, stand-alone software treatment planning system — Recall Details · AllClear