FDAJanuary 8, 2026device

Medline kits containing Tego Connectors Medline APHERESIS SUBQPORT ACCESS KIT SKU DYNDA2705 Medline CENTRAL LINE INSERTION SKU DYNJ63347A

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

What to do

FDA enforcement status: Ongoing

Brands named

medline industries lpmedlinemedline industries

UPCs

10193489468632401934894686331019348927806440193489278065

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.