FDAFebruary 20, 2020device

RayStation/RayPlan 9B, Build Number 9.1.0.933, stand-alone software treatment planning system

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

There is a potential for erroneous propagation of the current tabletop positions when applying pitch or roll correction. Even though pitch and roll correction is never propagated, non-zero pitch or roll causes the lateral, longitudinal and vertical values propagated to be incorrect.

What to do

FDA enforcement status: Terminated

Brands named

raysearch laboratories abraysearchraysearch laboratories

UPCs

07350002010266