FDAJune 22, 2022device

Crome Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D): a. Crome HF CRT-D, Model Numbers: DTPC2D4, DTPC2D1; b. Crome HF Quad CRT-D, Model Numbers: DTPC2QQ, DTPC2Q1

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

There is the potential for reduced shock energy (~79% of programmed energy) during high-voltage (HV) therapy for Cobalt and Crome implantable cardioverter defibrillators (ICDs) and cardiac resynchronization therapy defibrillators (CRT-Ds).

What to do

FDA enforcement status: Ongoing

Brands named

medtronic inc cardiac rhythm and heart failure crhfmedtronicmedtronic inc

UPCs

0076300017829100763000178314007630001783070076300017832100763000178253007630001782770076300017826000763000178284

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.