FDAApril 5, 2019device

Alinity i Processing Module, 03R65 01, with the below components. a. TED (Thermo Electric Device) Engine, Reagent cooler; b. CABLE, POWER OUT, TED CNTRLER Product Usage: The Alinity i processing module is a fully automated immunoassay analyzer allowing random and continuous access, as well as pri...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Potential loose cable connections on the reagent cooler, which could result in temperature errors and eventual reagent cooler failure. Could cause burn and shock injuries to users and a delay in test results. Device design has mitigated this risk, but it is a possibility.

What to do

FDA enforcement status: Terminated

Brands named

abbott gmbh co kgabbottabbott gmbh