FDAJanuary 5, 2026device

Raz-AT (Attendant Tilt) Mobile Shower Commode Chair, Catalog Number Z300 UDI-DI code: (01)00628758003007 Raz-ART (Attendant Rotational Tilt) Mobile Shower Commode Chair, Catalog Number Z333 UDI-DI code: (01)00628758003335 Raz-AT600 (Bariatric Tilt) Mobile Shower Commode Chair, Catalog Number Z360...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Due to a modification by a vendor and may be reinstalled without properly engaging the outside rear seat brackets that are used to keep the seat in place.

What to do

FDA enforcement status: Ongoing

Brands named

raz designraz

UPCs

00628758003007006287580033350062875800360100628758001003

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Raz-AT (Attendant Tilt) Mobile Shower Commode Chair, Catalog Number Z300 UDI-DI code: (01)00628758003007 Raz-ART (Attendant Rotational Tilt) Mobile Shower Commode Chair, Catalog Number Z333 UDI-DI code: (01)00628758003335 Raz-AT600 (Bariatric Tilt) Mobile Shower Commode Chair, Catalog Number Z360... — Recall Details · AllClear