FDASeptember 12, 2019device

Stryker, REF: 0250070460, 5.0 mm x 32 cm, Strykeprobe, Outer Sheath Replacement, RX Only, Non-Sterile, CE, UDI: (01) 07613327051971 These products are distributed with the following sheath/probe: 0250070441-PKG, StrykeFlow Electrocautery Probe, Spatula Tip 5MM 0250070442-PKG, StrykeFlow Electroca...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Potential diff. in the length of the sheath tube and the sheath/probe. A diff. in length could lead to the base of the sheath/probe not being fully seated in the sheath tube which could result in the distal tip of the sheath melting. This could lead to a compromised insulation, leading to unintentional flow of electricity to the patient. The sheath/probe is packaged separately.

What to do

FDA enforcement status: Terminated

Brands named

stryker

UPCs

07613327051971

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Stryker, REF: 0250070460, 5.0 mm x 32 cm, Strykeprobe, Outer Sheath Replacement, RX Only, Non-Sterile, CE, UDI: (01) 07613327051971 These products are distributed with the following sheath/probe: 0250070441-PKG, StrykeFlow Electrocautery Probe, Spatula Tip 5MM 0250070442-PKG, StrykeFlow Electroca... — Recall Details · AllClear