FDAMarch 13, 2024device

Patient Cart, Part: 380601, a component of the da Vinci SP Surgical System, Model: SP1098

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Surgical system may have inadequately welded top and/or bottom belts in the link 2 of the entry guide manipulator assembly/instrument arm joints, that may cause the following, 1) Top Failure: internal tissue injuries, body wall injury or emergent conversion to open surgery, 2) Bottom failure: pinching injury to the user with no harm to the patient.

What to do

FDA enforcement status: Ongoing

Brands named

intuitive surgicalintuitive