FDAJune 27, 2022device

HAMILTON-C6 Intensive Care Ventilator, REF: 160021

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Ventilator status indicator board can become loose, which could lead to water ingress (disinfectants) that may lead to technical fault alarms. Multiple technical faults in a short time may force ventilators into Safety Ventilation (blower runs constantly) or Ambient Sate (inspiratory channel/expiratory valves opened; patient breaths room air unassisted) with Panel connection lost message displayed

What to do

FDA enforcement status: Ongoing