FDAApril 6, 2021device

CareLink SmartSync Device Manager - Product Usage: used to treat arrhythmias related to bradycardia. Additionally, CRT-P devices can also provide cardiac resynchronization therapy (CRT) in treatment of heart failure, Model Number 24970A

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

A longevity estimation error may occur in the early years of device life when a unipolar pacing vector is programmed in the right atrial (RA) lead and/or the right ventricular (RV) lead. No other device features or therapies are impacted.

What to do

FDA enforcement status: Ongoing

Brands named

medtronic inc cardiac rhythm and heart failure crhfmedtronicmedtronic inc