Karl Storz RHINO-LARYNGO-BRONCHO FIBERSCOPE, Model # 11005BC1
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
During a complaint investigation it was determined that an error was observed within the instruction manual included with some scopes. Specifically, The Quick Connector Kits (QCK) and tray information, specified for the STERIS System 1E (SS1E) sterilization method within the instruction manual is incorrect. If the user uses SS1E to sterilize the impacted scopes and referred to the manual concerning the QCK information, the sterility of reprocessed scope cannot be guaranteed.
What to do
FDA enforcement status: Ongoing
Brands named
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- FDATangent Endoscopy, Tangent Single Use Digital Catheter. Model Numbers: (1) TNG4002-IND. (2) TNG-4007-6pk.2026-03-23
- FDAKARL STORZ - ENDOSKOPE, REF: 27000L,Uretero-Renoscope, 7 Fr., 6¿, 43 cm, NON STERILE, RxONLY, CE 01232024-12-19
- FDAKARL STORZ - ENDOSKOPE, REF: 27003L, MICHEL Uretero-Renoscope, 9.5 Fr., 43cm, RxONLY, CE01232024-12-19
- FDAKARL STORZ - ENDOSKOPE, REF: 27011K, Ureteroscope, 8 Fr., 6¿, 34cm, CE 01232024-12-19
- FDAKARL STORZ - ENDOSKOPE, REF: 27030KA, Cysto-Urethroscope, NON STERILE, CE2024-12-19
- FDAKARL STORZ - ENDOSKOPE, REF: 27002KP, Pediatric Uretero-Renoscope 7.3 Fr., NON STERILE, CE 01232024-12-19
- FDAKARL STORZ - ENDOSKOPE, REF: 27001L, Uretero-Renoscope, 8 Fr., 6¿, 43 cm, NON STERILE, RxONLY, CE 01232024-12-19
- FDAKARL STORZ - ENDOSKOPE, REF: 27002L, Uretero-Renoscope, 9.5 Fr., 6¿, 43 cm, NON STERILE, RxONLY, CE 01232024-12-19
Own something like this?
AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.
Start free — 200 items, no card →