FDAMarch 29, 2021device

VACUETTE¿ Blood Collection tube 9NC Coagulation sodium citrate 3.2% 13x75 blue cap-white ring, sandwich tube, non-ridged - Product Usage: indicated for use in collection and transport of venous whole blood for coagulation testing.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

VACUETTE¿ Blood Collection tube 9NC Coagulation sodium citrate 3.2% may show insufficient draw volume of the tubes, which could produce erroneous results for prothrombin time, activated partial thromboplastin time, and fibrinogen assays.

What to do

FDA enforcement status: Terminated

Brands named

greiner bio one north americagreinergreiner bio

UPCs

391200175744852912001757448839120017574607291200175746003912001757453929120017574532