AquaBplus and AquaBplus B2 Water Purification System - Product Usage: Intended for use with a hemodialysis system to remove organic and inorganic substances and microbial contaminants from the water used for hemodialysis or related therapies. Fresenius P/N 24-2500-0 All 3 units make up the AquaBp...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

AquaBplus HF module on the AquaBplus 2500 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2500 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation

What to do

FDA enforcement status: Terminated