FDAFebruary 13, 2020device

Dimension Vista¿ Total Bilirubin Flex¿ reagent cartridge -In vitro Diagnostic for the determination of total bilirubin. SMN: 10445146 Product Usage: Dimension Vista TBIL method is an in vitro diagnostic test intended to quantitatively measure total bilirubin in human serum and plasma on the Dimen...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Bilirubin assays may exhibit a positive bias with patients on eltrombopag therapy

What to do

FDA enforcement status: Terminated

Brands named

siemens healthcare diagnosticssiemenssiemens healthcare

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.