FDAApril 17, 2015device

Ultraview SL Command Modules, Model 91496-M, upgraded with the Masimo SpO2 Option. Product Usage: The Spacelabs Multi-parameter Module is intended for use with the Patient Care Management System (PCMS) to acquire, monitor, and process various clinical parameters from an adult or neonate/infant po...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Ultraview SL Command Modules which were upgraded with the Masimo SpO2 Option, Model 91496-M was affected. Customer reported that during the time that the module is resetting (~10 seconds), monitoring of all parameters will be suspended. Upon completion of this sequence, alarms will be reset to their default values.

What to do

FDA enforcement status: Terminated

Brands named

spacelabs healthcarespacelabs

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Ultraview SL Command Modules, Model 91496-M, upgraded with the Masimo SpO2 Option. Product Usage: The Spacelabs Multi-parameter Module is intended for use with the Patient Care Management System (PCMS) to acquire, monitor, and process various clinical parameters from an adult or neonate/infant po... — Recall Details · AllClear