FDADecember 19, 2023device

Smiths Medical Model 3010 Syringe pumps: a. Protege Model 3010, b. Medfusion Model 3010A, c. Protege UK Model 3010E

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Medfusion Model 4000, 3500, and 3010 pumps manufactured or serviced with Barrel Clamp Guide (Part Number G6000716), Lot Numbers P0407365, P0486670, and P0561740 may contain a molding defect that could potentially lead to slippage of the spring within the barrel clamp assembly. If this occurs, it could result in the inability of the pump to recognize a syringe or the pump may misidentify the size of the syringe loaded.

What to do

FDA enforcement status: Ongoing

Brands named

smiths medical asdsmithssmiths medical

UPCs

10610586032097

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Smiths Medical Model 3010 Syringe pumps: a. Protege Model 3010, b. Medfusion Model 3010A, c. Protege UK Model 3010E — Recall Details · AllClear