FDAMarch 20, 2025device
GenetiSure Dx Labeling Kit, REF: K1201-64105, contains human reference DNA female aliquot, REF5190-7317. A required kit for the GenetiSure Dx Postnatal assay (K1201A).
Official recall recorded by the U.S. Food & Drug Administration.
The hazard
DNA Labeling Kit for use in sample labeling to be hybridized to oligonucleotide microarrays has a component, Human Reference DNA Female aliquot, affected by a mosaic aberration that may result in Chromosome 12 baseline shift, which may lead to false positive/negative results, resulting in diagnosis delay, and may lead to delayed treatment of a treatable genetic disorder.
What to do
FDA enforcement status: Ongoing
Brands named
agilent technologiesagilent
UPCs
05700571112601
Recall history
Other federal recalls tied to the same brand — a factual pattern, straight from the source data.
- CPSCDaikin Comfort Technologies Manufacturing Recalls Amana Air Conditioners and Heat Pumps Due to Risk of Serious Injury from Fire and Burns2026-06-25
- CPSCAnalemma Water Bottles Recalled Due to Risk of Serious Injury or Death from Laceration and Ingestion Hazards; Imported by New Earth Technologies d.o.o.2026-05-07
- FDAPhaseOne Antimicrobial Solution, Model/Catalog Number: 150502026-04-27
- FDAGE HealthCare IOU ASSY ARK2250 CARESCAPE TELEMETRY SERVER, REF #2108729-901, Product Code SS9; ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)2026-04-24
- FDAGE HealthCare ApexPro v6.0 eDelivery Software, REF #5700026, Product Code SUV; used in conjunction wtih ApexPro CARESCAPE Telemetry Server (CTS) v6.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)2026-04-24
- FDAGE HealthCare CARESCAPE TELEMETRY SERVER ARK-2250L - NORTH AMERICA, REF #5697603, Product Code KSA; used in conjunction wtih ApexPro CARESCAPE Telemetry Server (CTS) v6.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)2026-04-24
- FDAGE HealthCare CARESCAPE TELEMETRY SERVER V5 ATO MODEL, REF #2063702-101, Product Code SS9; used in conjunction with ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)2026-04-24
- FDAGE HealthCare APEXPRO v5.0 - ARK2250 - Software Disaster Recovery USB Kit - Orderable by GE Service only, REF #2108523-01; ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)2026-04-24
Own something like this?