FDAMarch 20, 2025device

GenetiSure Dx Labeling Kit, REF: K1201-64105, contains human reference DNA female aliquot, REF5190-7317. A required kit for the GenetiSure Dx Postnatal assay (K1201A).

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

DNA Labeling Kit for use in sample labeling to be hybridized to oligonucleotide microarrays has a component, Human Reference DNA Female aliquot, affected by a mosaic aberration that may result in Chromosome 12 baseline shift, which may lead to false positive/negative results, resulting in diagnosis delay, and may lead to delayed treatment of a treatable genetic disorder.

What to do

FDA enforcement status: Ongoing

Brands named

agilent technologiesagilent

UPCs

05700571112601