FDAFebruary 6, 2026device

Owner's Booklets and Instructions for Use that are used with the following blood glucose measuring devices. Description/Item: Walgreens TRUE METRIX AIR Kit/REA4001-01, Leader TRUE METRIX AIR Kit/REA4002-01, CVS TRUE METRIX AIR Kit/REA4007-01, GNP TRUE METRIX AIR Kit/REA4011-01, Meijer TRUE METRIX...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The system labeling (user manual and online labeling) did not provide adequate directions for lay users to act on the E-5 error code. The error code description and recommended actions for the E-5 error code in the labeling do not: (a) clearly emphasize that an E-5 error code could represent a very high blood glucose level, and (b) prominently convey the appropriate immediate actions, such as contacting a healthcare professional. This could lead to a delay in treatment if the user does not seek medical attention immediately when receiving an E-5 error code and experiencing symptoms of high glucose.

What to do

FDA enforcement status: Ongoing

Brands named

trividia healthtrividia

UPCs

31191717389496295129267504283547668770142739870882068565941415072787933510227844122000916910939742445118225900512129201501521292007164

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.