FDAOctober 5, 2023device

MEDFLEX, REF: 1119-00; MEDBANK MAIN, REF: 1137-00; MEDFLEX 2.0, REF: 1139-00; MEDBANK TOWER MAIN, REF: 1145-00; MEDBANK MINI MAIN, REF: 1147-00 BD PYXIS MEDBANK Facility Software, REF: 139090-01; BD PYXS MEDBNK TWR MN 2HH1FH2HM2FM2FG1FF, REF: 169-100; BD PYXIS MEDBANK MINI 1HH-1FH- 1HM, REF: 169-...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

If "Clear All" selected medications is selected on patient profile, the automated dispensing cabinet may allow dispensing of medication without first requiring the RxVerify Validation Code be entered/without validation from a pharmacist, which could lead to unintended and incorrect medication removal.

What to do

FDA enforcement status: Ongoing

Brands named

carefusion 303carefusion

UPCs

1088540351270410885403517167108854035171811088540351256810885403512544

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
MEDFLEX, REF: 1119-00; MEDBANK MAIN, REF: 1137-00; MEDFLEX 2.0, REF: 1139-00; MEDBANK TOWER MAIN, REF: 1145-00; MEDBANK MINI MAIN, REF: 1147-00 BD PYXIS MEDBANK Facility Software, REF: 139090-01; BD PYXS MEDBNK TWR MN 2HH1FH2HM2FM2FG1FF, REF: 169-100; BD PYXIS MEDBANK MINI 1HH-1FH- 1HM, REF: 169-... — Recall Details · AllClear