PVT Interface Module (PVT)-Laboratory Automation system Brand names of the same product: FlexLab (FLX), Accelerator a3600 (ACP), Aptio Automation (AP2) Part Numbers: 72747000.A

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Firmware versions have the potential to mis-associate sample IDs leading to incorrect results or delayed sodium, potassium, and chloride patient results. Falsely increased or decreased electrolytes can lead to either inappropriate treatment of normal results, or failure to treat abnormal results resulting in abnormal levels. Abnormal potassium levels can result in weakness, polyuria, ileus, psychiatric disturbances, cardiac arrhythmias, respiratory depression, and death. Abnormal sodium levels can result in CNS disturbances and disturbances of water balance. Abnormal chloride levels can lead to acid base disturbances which can lead to respiratory and cardiac compromise. The event may occur only if all the following conditions occurs in few milliseconds timeframe: - The module is releasing a sample tube (Tube A) just placed into the carrier - Another sample tube (Tube B) is erroneously not diverted into the module buffer lane due to a malfunction of the divert gate Only in this

What to do

FDA enforcement status: Ongoing