FDAMarch 25, 2016device

Magellan Robotic Catheter Accessory Kits, Model number MACC 50 and MACC80. Cardiology: The Hansen Medical Magellan System is a robotic control system that enables positioning and visualization of a Magellan Catheter and an off-the-shelf guide wire while allowing a physician to remain seated and a...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Firm's quarterly dose audit testing showed that one batch was radiated below the augmented dose and therefore did not meet the augmented dose of 28.4 kGy.

What to do

FDA enforcement status: Terminated

Brands named

hansen medicalhansen

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Magellan Robotic Catheter Accessory Kits, Model number MACC 50 and MACC80. Cardiology: The Hansen Medical Magellan System is a robotic control system that enables positioning and visualization of a Magellan Catheter and an off-the-shelf guide wire while allowing a physician to remain seated and a... — Recall Details · AllClear