FDAFebruary 16, 2016device

Automated peritoneal dialysis (APD) cycler

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Baxter Healthcare Corporation is sending this communication to inform you of incomplete instructions in the AMIA Automated Peritoneal Dialysis (PO) System Clinician Guide for calculating the total recommended solution therapy volume. Specifically, the instructions do not specify the need for an extra 200 mL of PD solution in order to prime the patient line and for air purge operations. The tot

What to do

FDA enforcement status: Terminated

Brands named

baxter healthcarebaxter