FDASeptember 5, 2015device

ACUSON X700 Ultrasound System with software version 1.0.04. Model numbers: 10658844  ACUSON X700 Ultrasound System 10658846 - ACUSON X700 Ultrasound System (Russia) 10658845 - ACUSON X700 Ultrasound System (Korea) Radiology: The Siemens ACUSON X700 ultrasound imaging system is intended for the f...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Due to a communication error between the software and V5Ms transducers rotation function, ACUSON X700 ultrasound systems at software versions 1.1.04, display an IMG_15 error message causes the system to lock up, which requires a reboot to recover system operation.

What to do

FDA enforcement status: Terminated

Brands named

siemens medical solutionssiemenssiemens medical