FDAApril 10, 2015device

EPIQ 5 Ultrasound System with Pediatric Cardiology option, Model: EPIQ 5G, EPIC 5C, EPIQ 5W. Catalog Number: 795204 / 795205. Part Number: 989605408541 With 453561736761 (1.1) 453561750041 (1.1.1) 453561753651 (1.1.2) 453561772231 (1.2) 453561772611 (1.2.1) 453561786571 (1.2.2) 453561785081 (1.3)...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

When Epiq 5 Ultrasound System, WITH the Pediatric Cardiology option, is set up to Metric and weight and/or height is entered, a unit conversion error may result in the incorrect calculation of the patients Body Surface Area (BSA).

What to do

FDA enforcement status: Terminated