FDAFebruary 18, 2016device

The Livongo Health In Touch¿ Blood Glucose Monitoring System (BGMS), is an Over-The-Counter (OTC) system designed for the self- monitoring of blood glucose by persons with diabetes in home settings to aid in their diabetes management.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The instructional insert for the In Touch Control Solution for use with the In Touch meter and In Touch strips as a quality control check did not contain the limitation of sensor accuracy: (1) with the use of acetaminophen (Tylenol) (2) with the use of Ascorbic Acid (vitamin C) supplementation (3) in the presence of uric acid

What to do

FDA enforcement status: Terminated

Brands named

livongo healthlivongo

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.