FDAJanuary 23, 2026device

Volar Radius Plate with Angular Stability with the following description: Radius Plate PROlock II. Article Numbers: 21031-3, 21031-5, 21032-11, 21032-3, 21032-5, 21033-3, 21033-5, 21034-3, 21034-5, 21035-3, 21036-3.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

What to do

FDA enforcement status: Ongoing

Brands named

i t s

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.