FDAApril 13, 2016device

MAGNETOM Aera/Skyra Systems, whole body magnetic resonance imaging systems, Model numbers 10432914, 10432915. Docking station top cover (part no# 10591430) of the removable patient table.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The docking station top cover of the removable patient table may contain a magnetic metal plate. When docking/undocking the table, the magnetic top cover may become attracted to the magnet.

What to do

FDA enforcement status: Terminated