FDAFebruary 23, 2024device

ACUSON Maple Diagnostic Ultrasound System, REF: 11711750, ACUSON Maple Select, REF: 11711779, running software version VA10A

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

On ultrasound systems, when Cardiac DICOM SR feature is configured to display either minimum or maximum measured value, and multiple cardiac region measurements are made, and results are exported into the SR feature, then The SR viewer will display the LAST measured value, not min or max, which could contribute to patient condition misdiagnosis or negatively influence patient management decisions.

What to do

FDA enforcement status: Ongoing

Brands named

siemens medical solutionssiemenssiemens medical

UPCs

0405686999580904056869995830

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →