FDAJanuary 23, 2026device

I.T.S. Fibula Plate PROlock with Angular Stability with the below descriptions and corresponding article numbers. 1. Distal Tubular Plate; Article Numbers: 21602-3, 21602-4, 21602-5, 21602-6, 21602-7, 21602-8, 21603-3, 21603-4, 21603-5, 21603-6, 21603-7, 21603-8. 2. Fibula Plate 3.5mm; Article Nu...

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

What to do

FDA enforcement status: Ongoing

Brands named

i t s

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
I.T.S. Fibula Plate PROlock with Angular Stability with the below descriptions and corresponding article numbers. 1. Distal Tubular Plate; Article Numbers: 21602-3, 21602-4, 21602-5, 21602-6, 21602-7, 21602-8, 21603-3, 21603-4, 21603-5, 21603-6, 21603-7, 21603-8. 2. Fibula Plate 3.5mm; Article Nu... — Recall Details · AllClear