FDAMarch 26, 2018device

Mutlidiagnost-Eleva with Flat Detector 708034 As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The fixation of the upper and lower tilt actuator, of the MD-Eleva system might break off and the table will start to rotate from 0 to + 90 I- 90 degree with high speed. This rotating movement cannot be stopped by the user.

What to do

FDA enforcement status: Terminated

Brands named

philips medical systems nederlandsphilipsphilips medical

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Mutlidiagnost-Eleva with Flat Detector 708034 As a multifunctional universal imaging application system, General RJF, Fluoroscopy, Radiography and Angiography can be Performed along with more specialized interventional applications on human patients. — Recall Details · AllClear