FDAFebruary 28, 2024device

Mariner RDX MIS Polyaxial Head, REF: MX1-001010, part of Mariner RDX MIS Alpha Instruments and Implants, REF: AMRDXMIS, which are pedicle screw systems.

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Screw heads, of pedicle screw systems, have partial thread within the screw head that pose a risk of intraoperative cross-threading within the locking cap/screw head interface.

What to do

FDA enforcement status: Ongoing

Brands named

seaspine orthopedicsseaspine

UPCs

1088998129864710889981303297

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.

Start free — 200 items, no card →
Mariner RDX MIS Polyaxial Head, REF: MX1-001010, part of Mariner RDX MIS Alpha Instruments and Implants, REF: AMRDXMIS, which are pedicle screw systems. — Recall Details · AllClear