FDAApril 14, 2015device

MYLA server HP Proliant: DL380-G8 and ML350-G6 version 3.xx computer application software. MYLA V3 is a computer application ("Middleware") based on Web 2.0 technology which: Interfaces between the instruments connected to the application and the LIS(s) (Laboratory Information System(s).

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

The MYLA¿ server could slow down due to the volume (weight) of the data to manage and it could also stop processing, impacting the server start-up phase. The problem occurs in high volume settings. The MYLA server slows down and will stop processing completely if any task requires more than 5 minutes to complete. The server then fails to reboot and requires service, provided remotely or on-site

What to do

FDA enforcement status: Terminated

Brands named

biomerieux