FDAMarch 5, 2020device

Philips V60 Ventilators with Power Management PCBA part number 1055906

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

Solder connection failure on Power Management printed circuit board assembly of ventilators may cause 1) Blower to lose power, spool down, and trigger visual and audible High Priority Check Vent alarm indicating a need for alternative ventilation, or 2) Intermittent connection may disrupt expected operation and trigger unexpected shutdown without alarm. Failures may lead to hypercarbia/hypoxemia

What to do

FDA enforcement status: Ongoing

Brands named

respironics californiarespironics

Recall history

Other federal recalls tied to the same brand — a factual pattern, straight from the source data.

Own something like this?

AllClear screens your whole list against every CPSC, FDA, USDA and NHTSA recall — and tells you the day a new one names something you own.