FDAOctober 2, 2019device

Component for Laser Scanning Microscopes (LSM 700, LSM 800 and LSM 900)

Official recall recorded by the U.S. Food & Drug Administration.

The hazard

When the user tilts the transmitted light arm of the microscope to the end of its axis range, laser safety can be bypassed and result in emission of class lIlb laser radiation. This issue arises when a ZEISS inverted stage stand (Axio Observer or Axio Vert) is used in as a component of the LSM 700, 800 and 900 system

What to do

FDA enforcement status: Terminated

Brands named

zeiss carlzeisszeiss carl